AKD-A6 Automatic Karl Fischer Moisture Analyzer
The testing method of the AKD-A6 fully automatic Karl Fischer coulomb moisture analyzer from Aikorod brand is the Karl Fischer coulomb method, which i
Product details



Acored brandThe testing method of AKD-A6 fully automatic Karl Fischer moisture analyzer is Karl Fischer Coulomb method, which is a globally recognized method for testing water content. The instrument adopts a powerful new generation processor and dual CPU design, greatly improving computational efficiency. The design of dual power channels separates the electrolytic electrode from the measuring electrode, and the instrument can automatically remove various interferences, greatly improving the accuracy of test results, which can reach 0.0001% (1ppm). Built in multiple calculation formulas to meet the testing needs of different samples, capable of storing 1000 experimental data, and equipped with a printer to directly print out experimental data.
Acored brandAKD-A6 fully automatic Karl Fischer moisture analyzerIt has the advantages of high measurement accuracy, good repeatability, and low testing cost, and is widely used in industries such as petroleum, chemical, pharmaceutical, power, scientific research, and education. It can be used to measure trace amounts of moisture in various liquids, solids, and gases.

Acored brandAKD-A6 fully automatic Karl Fischer moisture analyzerTechnical parameters:
model: | AKD-A6 |
Measurement method: | Karl Fischer Coulomb coulometric method |
Display measurement range: | 0.01ug water -200mg water |
Water content range: | 0.0001%(1ppm)—100% |
resolving power: | 0.01ug water |
Display mode: | 10 inch 260000 color LCD touch screen display (800 * 600 pixels) |
Accuracy: |
When the water content is between 2ug and 1000ug, the measurement error is ≤± 2ug When the water content is above 1000ug of water, the measurement error is ≤± 0.2% (excluding injection error) |
Endpoint indication: | Screen display \ \ Sound warning \ \ Print output \ \ End light prompt |
Blank processing: | The blank current microprocessor automatically controls compensation to ensure accurate deduction of blank during the 10 minute sample enrichment time |
Mixing speed: | Adjustable control |
Drift compensation: | Microprocessor automatic control |
Data input: | Touch screen |
Sample Number: | User |
Electrolysis speed: | ≤ 2.4 mg water/min |
Electrolytic current: | 0-430mA |
Data storage: | 1000 items |
Printing function: | Built in thermal printer, 56mm paper width |
Print content: | Ug \ \ ppm \ mg/L \ \ Sample number \ \ Experimenter \ \ Date |
External functions: | Can connect to a computer and work both online and independently |
Self checking function: | Automatic diagnosis of instrument faults |
Calendar/Clock: | Analyze time and date display and print output (without losing power) |
External dimensions: | 430*350*190mm |
Power consumption: | Less than 40W |
Using power supply: | Voltage 220V ± 10%, frequency 50HZ ± 2.5HZ |
Usage environment: | Temperature 5-40 ℃, humidity<85% |
Weight: | About 7.5KG |

Acored brandAKD-A6 fully automatic Karl Fischer moisture analyzerScope of application:
liquid |
Pharmaceutical industry: raw materials for traditional Chinese and Western medicine, etc Pesticide industry: emulsification, etc Petroleum and power industry: insulation oil, transformer oil, etc Chemical industry: acids, alcohols, benzenes, phenols, ethers, organic solvents, etc Other industries: lithium batteries, electrolytes, etc |
solid | Solid substances with good solubility, such as various inorganic salts and citric acid |
gas | Natural gas, liquefied gas, fluorocarbons, butadiene and other gases |

Acored brandAKD-A6 fully automatic Karl Fischer moisture analyzerCompliant with standards:
GB/T 6283-2008 Determination of Moisture Content in Chemical Products (Karl Fischer Method)
ASTM E1064-2008 Karl Fischer Coulombic Titration for Determination of Water Content in Organic Liquids
GB/T 7600-1987 Determination of Moisture Content in Operating Transformer Oil (Coulombic Method)
ASTM D4928-00 (2010) Karl Fischer Coulombic Titration for Determination of Water Content in Crude Oil
ASTM D6304-2007 Karl Fischer coulometric titration method for determining water content in petroleum products, lubricants, and additives
ISO 10337-1997 Determination of Moisture in Crude Oil (Karl Fischer Coulombic Titration Method)
GB/T 11146-2009 Determination of Crude Oil Water Content (Karl Fischer Coulombic Titration Method)
GB/T 3727-2003 Determination of Trace Water in Ethylene and Propylene for Industrial Use
GB/T 5074-1985 Method for determination of moisture content in coking products by microcoulometry
GB/T 6023-2008 Determination of Trace Water in Industrial Butadiene (Karl Fischer Coulombic Method)
GB/T 7376-2008 Determination of Trace Moisture in Fluoroalkanes for Industrial Use (Karl Fischer Method)
GB/T 18619.1-2002 Determination of Water Content in Natural Gas (Karl Fischer Coulombic Method)
GB/T 18826-2002 Industrial 1,1,1,2-tetrafluoroethane HFC-134a
SH/T 0246-1992 Determination of Water Content in Light Petroleum Products (Electricity Method)
SH/T 0255-1992 Method for determination of moisture content in additives and lubricating oils containing additives (electric method)
ASTM E1064-2008 Karl Fischer Coulombic Titration for Determination of Water Content in Organic Liquids
ASTM D4928-00 (2010) Karl Fischer Coulombic Titration for Determination of Water Content in Crude Oil
ASTM D6304-2007 Standard for Determination of Water Content in Petroleum Products, Lubricants, and Additives by Karl Fischer Coulombic Titration

Acored brandAKD-A6 fully automatic Karl Fischer moisture analyzerHas the following characteristics:
1. 10 inch 260000 color touch screen display, high-energy microprocessor, real-time display of instrument working status;
2. Adopting a composite control system of microcontroller and computer, dual CPU design, intelligent analysis and determination of moisture content;
3. The dual power channel separates the electrolytic electrode from the measuring electrode, and the instrument can automatically control various interferences, greatly improving the accuracy of test results;
4. Built in multiple calculation formulas, users can choose the appropriate testing method, and the instrument will automatically calculate the water score based on the selected formula;
5. The instrument automatically measures the sample, analyzes the results and displays them, and can automatically calculate the average value of multiple measurements of the sample;
6. Automatically calculate and print content such as ug, ppm, percentage content, mg/L, sample name, and experimental date;
7Automatic data storage, capable of storing 1000 sets of data, convenient for viewing historical records;
8. Multiple language display operation, instrument program can be upgraded when national standards are revised;
9. The blank current microprocessor automatically controls compensation, and the reagent can quickly reach equilibrium state;
10. The instrument has a self checking function. If there is a short circuit or open circuit fault in the electrode, the instrument will automatically prompt the user;
11. 0-430mA high electrolysis current, high detection sensitivity and fast analysis speed;
12. Streamlined appearance design, surface electrostatic spray treatment, anti-corrosion, easy to clean;
13. The perpetual calendar function includes year, month, day, week, hour, minute, and second, which are automatically stored after power failure and will not be lost for 10 years.

Acored brandAKD-A6 fully automatic Karl Fischer moisture analyzerOperation steps:
1. After installing the instrument accessories according to the instructions, turn on the instrument;
2. Set parameters according to the instructions, select different setting items, and press the confirm button to save after selection;
3. Reagent equilibrium. Inject an appropriate amount of pure water into a 50 microliter sampler to neutralize any excess water in the reagent;
4. Instrument calibration. Using a 0.5ul sampler to extract 0.1ul of pure water for injection, the displayed result should be "100 ± 10ug (excluding injection error)". If the displayed result is within the error range, sample measurement can be performed.
5. Sample determination. After pressing the start button to inject the sample, the instrument will automatically perform the test. After the test is completed, a beep alarm will sound and the display screen will indicate "Test completed".

Acored brandAKD-A6 fully automatic Karl Fischer moisture analyzermatters needing attention:
1、 Cannot display normally
Please check if the power connection wire, fuse, and power switch of the instrument are functioning properly.
2、 The instrument displays iodine
1. Evaluate whether the reagent is over iodine. If so, use a 0.5 microliter sampler to extract 0.2-0.4 microliters of water and inject it. Do not use an injector of 50 microliters or larger for injection.
2. Check whether the platinum wire at the bottom of the measuring electrode is connected together, causing a short circuit.
3、 The instrument shows an open circuit
Check if the measuring electrode is in good contact and re insert it firmly. Check if the connecting wire is broken.
4、 Open electrolysis without counting
Check if the electrolytic electrode is in good contact, unplug and plug it again, and check if the connecting wire is broken.
5、 Electrolysis does not end
Check if the reagent has expired.